Summary
The confirmed rate counts tenders with at least one human-reviewed, confirmed integrity finding. It is a risk signal, not an accusation -- most flagged tenders have not been reviewed yet.
Authority facts
Integrity findings
5Automated screening based on procurement documents and award history. Findings are only shown here once a human reviewer has confirmed them; this is not an allegation of wrongdoing.
Specification: 5
-
Specification
· High
· Laparoskopisko operāciju video sistēma
The requirement for a specific number of 12G-SDI outputs is challenged by the market as an obsolete technology for 4K systems, where HDMI or DisplayPort are industry standards. By mandating a specific, outdated interface count, the specification narrows the field to legacy equipment, effectively excluding modern systems that utilize standard 4K connectivity.
-
Specification
· High
· Laparoskopisko operāciju video sistēma
The requirement mandates 'continuous' auto-focus and manual focus via buttons. Industry feedback indicates that continuous auto-focus is often clinically unsuitable for laparoscopy due to focus hunting, and button-based manual focus is ergonomically inferior to ring-based focus. This combination of specific, non-standard operational requirements appears tailored to a specific manufacturer's control logic.
-
Specification
· High
· Laparoskopisko operāciju video sistēma
The requirement for 'at least two different speeds' for automatic smoke evacuation is identified as technically unjustified, as modern systems typically use a single optimized flow rate. Mandating multiple specific speeds restricts the procurement to systems designed with this specific, non-essential feature, excluding other effective, modern solutions.
-
Specification
· High
· Bioķīmijas, Imūnķīmijas, Hematoloģijas un Koaguloģijas analizatoru lietošanas tiesību piešķiršana par atbilstošu reaģentu iegādi un medicīnisko ierīču tehniskā apkope lietošanas periodā
The authority mandates a maximum water consumption of 5 L/h for a biochemistry analyzer. Potential bidders have pointed out that this is incompatible with the required test volume and that modern analyzers typically require up to 20 L/h. By setting an artificially low ceiling on water consumption, the authority effectively excludes most modern, high-throughput analyzers that would otherwise meet the clinical requirements, likely tailoring the specification to a specific model that operates with lower water usage.
-
Specification
· High
· Bioķīmijas, Imūnķīmijas, Hematoloģijas un Koaguloģijas analizatoru lietošanas tiesību piešķiršana par atbilstošu reaģentu iegādi un medicīnisko ierīču tehniskā apkope lietošanas periodā
The requirement mandates the use of disposable reaction cuvettes with a specific capacity of at least 10,000 tests. This excludes systems that use reusable cuvettes (which are standard in many high-end biochemistry analyzers), thereby narrowing the competition to only those manufacturers that utilize a specific disposable cuvette system.
Recurring winners
10| Supplier | Awards | Total value |
|---|---|---|
| SIA "A.MEDICAL" | 8 | 166,963.50 EUR |
| Sabiedrība ar ierobežotu atbildību "ELME MESSER L" | 2 | 157,107.99 EUR |
| "TRADINTEK" SIA | 2 | 54,992.00 EUR |
| SIA "BVU" | 1 | 173,467.00 EUR |
| Sabiedrība ar ierobežotu atbildību "LATMED-PRO" | 1 | 76,225.90 EUR |
| SIA "VIRŠI Renergy" | 1 | 74,443.73 EUR |
| Sabiedrība ar ierobežotu atbildību "AOX Trade" | 1 | 71,504.92 EUR |
| SIA "OneMed" | 1 | 63,580.61 EUR |
| SIA “”PP”, Medicīnas centrs” | 1 | 50,179.00 EUR |
| Sabiedrība ar ierobežotu atbildību "DM Premium" | 1 | 37,566.35 EUR |