Summary
Integrity signals
30 · MediumAutomated screening based on procurement documents and award history. Findings are only shown here once a human reviewer has confirmed them; this is not an allegation of wrongdoing.
-
Specification
· High
The requirement for 'Plasmapore' (a specific proprietary surface coating technology) is used repeatedly across multiple items (2.1.1.1, 2.2.1.1, 2.3.1.1, 2.2.3.1, 2.3.3.1). Plasmapore is a specific trademarked surface technology of Aesculap/B. Braun, and mandating this specific coating excludes other manufacturers who use different, equally effective porous coating technologies (e.g., plasma spray, hydroxyapatite, or trabecular metal structures).
-
Pārklāta ar plazmoporu proksimālajā daļā.
-
-
Specification
· High
The specification mandates 'Isocer' ceramic for the femoral head (2.2.2.1, 2.3.2.1). Isocer is a specific proprietary brand name for a ceramic material used by a single manufacturer. Requiring a specific brand of ceramic material rather than specifying the required material properties (e.g., alumina-toughened zirconia or similar ISO standards) restricts competition to only one manufacturer.
-
Sastāv no Isocer keramikas.
-
Procurement facts
Values and statistics
Authority
No award data available.
Lots
1 lot| Lot | Place and duration | Estimated value | Result |
|---|---|---|---|
| Lot 1 | LVA | 200,000.00 EUR | Not provided |
Documents
1 document| Document | Type and metadata | Link |
|---|---|---|
| Procurement documents | documents_url | Open document |
Notice history
1 notice| Notice | Type and form | Date |
|---|---|---|
| 90c6dad9-775e-4c8b-9c76-d93ac89ebcdf | pil-prior-information | 2024-06-05 |